TheemoFisher TPXEVOS 恒温台
ThermoFisherTPXEVOS恒温台
The ThermoPlate™ System from Tokai Hit provides a fast, precise, and convenient system to maintain the temperature on the microscope stage of EVOS™ FL and XL imaging systems. The system affords accurate and reliable thermal control of specimens during observation on the microscope stage. A controlled and stable temperature is particularly important for sensitive cell types like oocytes, stem cells, and primary cells and allows extended inspection on the microscope stage without adverse cell viability impact.
The Tokai Hit ThermoPlate System offers you these advantages:
• Accurate and reliable thermal control of microscope stage
• Enhanced viability of sensitive cell types like oocytes and stem cells
• Thin, optical quality glass allows use with range of objectives and high-quality images
• Sterilizable sensor allows sample temperature measurements
• Compatible with EVOS FL and XL imaging systems
A sterilizable temperature sensor enables continuous temperature recording of both sample and surface temperatures. The temperature recordings are logged and saved in a CSV format that can be exported. Importantly the temperature measurement log can be used for quality control purposes.
The ThermoPlate System consists of a control unit and a strengthened optical-quality heated glass plate that can be maintained at up to 50°C, with a lower, reliable limit 5°C above ambient. A continuous current control method ensures a highly stable temperature control, with an accuracy of +/- 0.1°C. This prevents thermal drift that may reduce image quality over time-dependent observations at higher magnifications. The optical-quality glass causes minimal loss of light and allows the capture of high-quality images in both transmitted light and fluorescence. With a thickness of only 0.5 mm, a range of objectives for the EVOS imaging systems can be used to image samples on the ThermoPlate.
The EVOS stage plate (Cat. No. AMEP4684) was designed to hold the ThermoPlate and is required for its use on EVOS FL or XL imaging systems. Explore other vessel holders and stage plates for a perfect fit of your microscope slide, cell culture flask or dish, or microwell plate to the stage of the EVOS FL and XL imaging systems.
Explore the entire EVOS line of imaging systems and accessories.
System specifications
Heating plate
Frame (135 x 88 mm) with heating wire socket and cell vessel clamps
Strengthened glass plate (0.5 mm thickness)
Control unit
Temperature control method: PID control
Temperature increments: 0.1°C
Setting method: Digital switch with UP and DOWN keys
Max. temp. setting: 50°C
Min. temp. setting: Ambient plus 5°C
Temperature accuracy: +/- 0.3°C at heating plate surface, plate center
Temperature sensor: thermocouple
Other
For use with 120V to 240V power
Use location: indoor only
Relative humidity: 35% to 80% RH, non-condensing
TheemoFisher TPXEVOS 恒温台,吸附功能加热台
DywjPlate 加热体设计有气孔,可吸附薄片样品方便观察测量。
加热体温度控制室温~400℃,最大升温速度10℃/分钟,最慢升温速度0.1℃/分钟,系统波动度±0.5℃。
加热台有四个安装孔位,方便固定。
加热体无氧化、寿命长、体积小,高、低温有效隔离。